High-end modular Design Vital Sign Patient Monitor
这里是标题一h1占位文字
Classification :
Features
●15" Color TFT LCD Screen, optional touch screen, Display Resolution: 1024* 768
●2 hours wave storage,720 hours Trend review,200 alarm events, 1000 NIBP measurement data review
●Standard interface: Maximum 9-channel waveforms display, the waveform color can be defined by user (7 colors)
●Optional interface: Maximum 14-channel waveforms display, the waveform color can be defined by user (7 colors)
●ST and 13 kinds arrhythmia analysis, pacemaker analysis, drug dose calculation
●Respiration monitoring with apnea alarms, alarm review, arrhythmia review
●Unique human voice alarm, Three Level: Low, medium and high,
Indication: Auditory and visual, All alarms can be silenced, Alarm Volume: The scope is 1~10
●Rechargeable high energy buil-in Li battery, AC/DC available, silent fanless design
●Anti - defibrillation design, NIBP over pressure protection
●Central station support Nurse Call by wire/wireless network
●Digital SpO2 with excellent performance
●Support more than 20 kinds languages
●PVI (Pleth Variability Index)
●USB support PDF file export
●The machine includes a spare parts box
Technical Specifications
Quality Standards and Classification: ISO13485,ISO9001,CFDA,FSC
MDD classification: Class II b
Anti-electroshock degree: Class I equipment
(internal power supply)
TEMP/SpO2 /NIBP: BF
ECG/Resp: CF
Application Range
Adult/Pediatric/Neonatal/Medicine/Surgery/
Operating Room/ICU/CCU/Clinic/Ambulance/
Emergency Centers/Bedside Care Ward/Family Care
Environment
Temperature
Working:0 ~ 40℃
Storage: -20 ~ 50℃
Humidity
Working: 15% - 90%
Storage: 15% - 90% (no coagulation)
Atmospheric pressure
Working: 86.0kPa ~ 106.0kPa
Storage: 86.0kPa ~ 106.0kPa
Power Requirement
External power required
AC: 100V ~ 240V, 50Hz/60Hz
DC: Built-in rechargeable battery
Battery: 14.8V, 2200mAh rechargeable Li-ion battery
2h operating after full charge
5min operating after low battery alarm
Input power: Pmax=90VA
FUSE: T1.5AL 250V
Dimension and Weight
Equipment: 350mm x 155mm x320mm(about); 4.5kg(about)
Package: 430mm x 315mm x 400mm; 7.7kg(about)
Storage
Trend 720 hours
NIBP review 1000 NIBP events
Wave review 2 hours
Alarm review 200 alarm events
All storage data are non volatile
ECG
5 Leads: RA, LA,LL, RL, V
lead mode: I, I, II, AVR, AVL, AVF, V
3Leads:R,L,F or RA,LA, LL
Lead mode: I, II, II
Gain: 2.5mm/mV, 5.0mm/mV, 10mm/mV,20mm/mV, auto
Sweep speed: 12 .5mm/s, 25mm/s, 50mm/s
Bandwidth
Surgery: 1 ~ 20Hz
Monitor: 0.5 ~ 40Hz
Diagnostic: 0.05 ~ 130Hz
Calibration Signal: 1(mVp-p), Accuracy: 5%
CMRR
Diagnostic Mode: >90dB
Monitor Mode: > 110dB
Surgery Mode: >110dB
Electrode offset potential: ±300mV
Leakage Current: <10 uA
HR Measure Range
Adult: 15 ~ 300 bpm
Neonatal/Pediatric: 15 ~ 350 bpm
Accuracy: ±1%
Resolution: 1bpm
Sensitivity: >200 (uVp-p)
Differential Input Impedance: >5 MΩ
ST Segment
ST Segment Monitoring Range
Measure and Alarm: -2.0 ~ +2.0mV
Accuracy: -0.8mV ~ +0.8mV
±0.02mV or ±10%,which is greater
ARR Detecting
Type: ASYSTOLE, VFIBNTAC, COUPLET, BIGEMINY,
TRIGEMINY, R ON T, VT>2, BRADY, MISSED BEATS,
PNP, PNC, Alarm Available, Review Available
NIBP
Method: Oscillometry
Mode: Manual, Auto, STAT
Measuring Interval in AUTO Mode: 1,2,3, 4,5, 10, 15,
30, 60, 90,120, 180,
240, 480 (Min)
Measuring Period in STAT Mode: 5 Min
Pulse Rate Range: 40 ~ 240bpm
Alarm Type: SYS, DIA, MEAN
Measuring and alarm range:
Adult Mode: SYS 40 ~ 280mmHg
DIA 10 ~ 220mmHg
MEAN 20 ~ 240mmHg
Pediatric Mode: SYS 40 ~ 220mmHg
DIA 10 ~ 160mmHg
MEAN 20 ~ 170mmHg
Neonatal Mode: SYS 40 ~ 135mmHg
DIA 10 ~ 100mmHg
MEAN 20 ~ 110mmHg
Resolution: 1mmHg
Accuracy: Maximum Mean error ±5mmHg
Maximum Standard deviation 8mmHg
Overpressure Protection
Adult Mode: 300mmHg
Pediatric Mode: 240mmHg
Neonatal Mode: 150mmHg
Static pressure accuracy: 3mmHg
SpO2
Measuring Range: 0 ~ 100%
Resolution: 1%
Alarm Range: 0 ~ 100%
Accuracy: 70% ~ 100% ±2%
<69% unspecified
Actualization interval about: 1 Sec.
Alarm Delay: 10 Sec
Pulse Rate
Measuring and Alarm Range: 20 ~ 250bpm
Resolution: 1bpm
Accuracy: ±3bpm
Perfusion Index
Measure and Alarm Range: 0.02 ~ 20%
Resolution: 0.01%
Respiration
Method: Impedance between R-F(RA-LL)
Measuring Impedance Range: 0.3~3Ω
Base line Impedance Range: 200~4000Ω
Bandwidth: 0.3~2.5Hz
Resp rate: Measurement Range
Adult: 0~ 120 BrP
Neonatal / Pediatric: 0 ~ 150 BrPM
Resolution: 1 BrPM
Accuracy: 0 ~ 6 BrPM: unspecified
7 ~ 150 BrPM: ±2 BrPM
Apnea Alarm: 10 ~ 40 S
Gain 0.25,0.500,1,2,4
Differential Input Impedance: >2.5MΩ
Inspiriting current: <300 μA RMS max
TEMP
Channel: 2
Measuring and Alarm Range: 0 ~ 50℃
Resolution: 0.1℃
Accuracy: ±0.1℃ (0 ~ 50℃)
Actualization intervalabout: 1 Sec
Average Time Constant: <10 Sec
Sensor Type: 10K series, 2.25K series
Standard Accessories Adult
NIBP cuff & Extension tube, ECG cable & electrodes,
SpO2 sensor, TEMP probe, Lithium-ion battery,
Power cable,User manual
Technical Specifications
IBP (optional)
Sensor
Reusable sensors: OHMEDA P23XL or BD, EDWARD compatible
Once usage: OHMEDA DT-4812 or BD, EDWARD compatible
Excitation voltage: +5Vdc ±2%
Sensitivity: 5uVN/mmHg
Channels: 2 channels
Measure Range: -50~ 360(mmHg)
Resolution: 1mmHg
Accuracy(no sensor): ±2% or ±1mmHg, use the greater
Band Width
Normal mode: DC ~ 40Hz
Smooth mode: DC ~ 12.5Hz
Measurable parameter: ART, PA, CVP, LAP, RAP, ICP
CO2(optional)
Measure Range: 0-150mmHg
Resolution: 1mmHg
Accuracy: 2mmHg @ < 5.0% CO2(at ATPS)
Breath Rate: 3 ~ 150 bpm
Recorder (optional)
Paper width: 50 mm
Paper length: 20 m
Speed: 6. 25/12.5/25/50 mm/s
Wave channel: 3 channels
Gas (optional Sidestream EtCO2)
RangeFiCO2,EtCO2:0-10 VOL% 10-15VOL%
RR: 0~ 0-150BPM
N20:0-100VOL%
Sevoflurane: 0-10 VOL%
Isoflurane: 0-8 VOL%
Enflurane: 0-8 VOL%
Desflurane: 0-22 VOL%
Halflurane: 0-8 VOL%
Resp: 3- 120 BPM,1BPM
O2(Optional): 0~100
Accuracy
FiCO2,EtCO2: + (0.3% +2% of reading)@(0 ~ 10%)+ (0.3% + 5% of reading)@(0 ~ 10%)
RR
N20:+1BPM:+ (2% + 5% of reading)
Sevoflurane : (0.3% + 2% of reading)@(5 ~ 10%):+ (0.3% + 2% of reading)
Desflurane
Isofurane
:+ (0.3% + 2% of reading)@(0 ~ 5%)
Enflurane: I (0.3% + 2% of reading)@(5 ~ 8%)
Halflurane:+ (0.3% + 2% of reading)
Standard:3/5-leads ECG、SpO2、NIBP、RESP、2-Temp、PR/HR
Optional:WIFI、AG、2-IBP、VGA、CMS、EtCO2、Nurse Call、Suntech Blood Pressure、
Touch Screen、Nellcor Spo2、Wall Mount、Recorder、Rolling Stand、Depth of anesthesia、Children Accessories、Newborn Accessories、Disposable Spo2 Probe