Differences Between High-Frequency and Power-Frequency X-ray Machines
Release time:
2026-07-28
The use of energy-storage components greatly reduces mains-power demands; in many cases the system can run directly from a standard supply (large power-frequency units often require a dedicated feed). At the same time, the high-voltage transformer can be made much smaller, facilitating overall miniaturization of high-power generators.
High-frequency unit
- Generator switching frequency > 20 kHz.
- After rectification the high voltage is almost pure DC; ripple is < 0.1 %.
- Produces an almost mono-energetic electron beam → narrow, clean X-ray spectrum.
- Reduced scatter, 50 % lower total beam output for the same image quality, and noticeably sharper images.
Power-frequency (line-frequency) unit
- Generator frequency < 400 Hz (50/60 Hz mains).
- After rectification the high voltage still carries 100 Hz sinusoidal ripple; even after filtering the ripple remains > 10 %.
- Generates a broad X-ray spectrum with pronounced low-energy tails → more scattered radiation.
- Lower characteristic-line yield per unit time → poorer subject contrast and hazier images.
Since the 1990s all imported X-ray systems have been high-frequency; the international trend is clear and China is transitioning in the same direction.
Imported versus Domestic Molybdenum-Target X-ray Tubes
Domestic tube (custom order)
- Stationary (fixed) anode, single 0.5 mm focal spot.
- Lower heat capacity.
Imported tube
- Rotating anode, dual-focus 0.3 mm / 0.1 mm.
- Higher heat capacity.
- Routine radiography uses the 0.3 mm focus—its area is 1.5× smaller than that of the domestic tube.
- Smaller focal spot → sharper images, but manufacturing difficulty drives the cost up ~30× versus domestic tubes.
Measured performance
The domestic tube currently delivers 80 % of the spatial resolution of the imported tube, achieving 80 lp cm⁻¹—sufficient for all clinical requirements. Models B and D in our line use the imported tube and match the performance specifications of imported systems.
In terms of exposure-time accuracy, power-frequency X-ray units exhibit wide variation depending on the rectification method used, whereas high-frequency generators incorporate voltage pre-setting and closed-loop feedback, yielding far tighter control. This markedly reduces the soft-radiation component of the beam, lowers patient dose, and simultaneously improves image quality and overall X-ray output efficiency.

Structurally, high-frequency X-ray units are comparatively simple.
Because the kV and mA values are preset and regulated independently through purely electronic feedback, mechanical knobs, switches, and traditional kV-compensation or space-charge-compensation circuits are eliminated. This yields faster response, higher output accuracy, and allows kV and mA to be adjusted during the exposure itself.
The use of energy-storage components greatly reduces mains-power demands; in many cases the system can run directly from a standard supply (large power-frequency units often require a dedicated feed). At the same time, the high-voltage transformer can be made much smaller, facilitating overall miniaturization of high-power generators.
Latest News
A beam of 2.1-micron light is rewriting the history of the treatment for "stones and hyperplasia".
Lithotripsy is recognized as the "gold standard". Whether it is hard oxalate calcium stones, tough cystine stones, or oddly shaped antler-shaped stones, holmium laser can efficiently crush them; combined with the Moses pulse modulation technology, the displacement of stones is smaller and the efficiency is higher. It can even "use one light for multiple purposes" - during lithotripsy, it cuts open the inflammatory polyps surrounding the stones, treats ureteral stenosis, and solves multiple problems in one surgery.