• YH-6100 Anesthesia Machine
YH-6100 Anesthesia Machine

YH-6100 Anesthesia Machine

这里是标题一h1占位文字

The Anesthesia machine makes a good performance in Intensive Care Units (ICU), Operation room, Anesthesiology Department and other departments. Professional design for adult, child and infant inhalation anesthesia and respiratory management, with advanced ventilation modes. Outstanding ergonomic design, it ranks high level in safety, stability and convenience as well as user experiences.

Applications:

The Anesthesia machine makes a good performance in Intensive Care Units (ICU), Operation room, Anesthesiology Department and other departments.

Professional design for adult, child and infant inhalation anesthesia and respiratory management, with advanced ventilation modes. Outstanding ergonomic design, it ranks high level in safety, stability and convenience as well as user experiences.

This comfortable classic model, easy to use and be designed together with experienced experts to streamline your anesthesia workflow.

Trust point:

1.Simplicity: 4 static casters with self-locking function.

2.Precision in an anesthesia ventilator with multiple ventilation modes: IPPV, A/C, PCV, SIMV, SIGH and MANUAL.

3.Flexible configurations able to customize your requirements.

4.International standard and advanced technology suitable for wide range use.

5.Compact interface and big screen give you better operating experience.

6.Over 5,000 units are installed in more than 200 countries.

7.Designed and manufactured by our team with over 25 years of experience in ICU field.

Features:

1.10.4” TFT LCD screen displays the Ventilation parameters, Alarm information and Oscillogram.

2.High precision flow meter, instantly know the fresh gas flow to your patient.

3.Integrated breathing circuit design, ensure easy operating and keep tidy.

4.Multiple working modes such as volume control and pressure limit, adapt to wide range patient.

5.Vaporizer with temperature, pressure, flow compensation and self-lock function, keep safety anytime.

6.Multiple parameters monitoring interface, make every parameter clear, let users know the patient conditions in all aspects.

7.Pressure-time, low-time loop oscillogram and high precision ETCO2, O2 concentration show in real time.

8.Vital sign monitor and anesthetic gas monitor are optional.

9.ETCO2 module and Anesthesia Gas Scavenging System (AGSS) are optional.

Safety:

1.Three level alarm system, visual and sound alarm information.

2.With multiple type of alarm, reminder and protection functions.

3.Advanced power management control technology.

4.Built-in backup battery provide the emergency power supply to the unit.

5.Low O2pressure alarm and N2O cut-off protection.

Ventilation Mode

IPPV, A/C, PCV, SIMV, SIGH, MANUAL

Ventilator Parameter Range

Flowmeter

O2   (0.1~10 L/min)

N2O   (0.1~10 L/min)

AIR   (0.1~10 L/min)

Rapid oxygen supply   25 L/min~75 L/min

Tidal volume(Vt)   0, 20 mL~1500 mL

Frequence   1 /min~100 /min

I:E   4:1~1:8

PEEP   0 cmH2O~30 cmH2O

Pressure triggering sensitivity (Ptr)   -20 cmH2O~0 cmH2O (Based on PEEP)

Flow trigger sensitivity(Ftr)   0.5 L/min~30 L/min

Pressure control (PC)   5 cmH2O~60 cmH2O

SIGH   0 (off) 1/100~5/100

Apnea ventilation   OFF,5s~60s

Pressure limit   20 cmH2O~100 cmH2O

Monitoring Parameter

Frequence(Freq) 0 /min~100 /min

Tidal volume(Vt) 0 mL~2000 mL

MV 0 L/min~100 L/min

Oxygen concentration 15 %~100 %

Alarm and protection

The AC power failure alarm   Power failure or no connection

Low voltage alarm for back up battery   <11.3±0.3V

No tidal volume   ≤5mL within 6s

High oxygen concentration alarm   19%~100%

Low oxygen concentration alarm   18%~99%

High airway pressure alarm   20 cmH2O~100 cmH2O

Low airway pressure alarm   0 cmH2O~20 cmH2O

High minute volume alarm: 

Adult (5 L/min~20 L/min)

Paed (1 L/min~15 L/min)

Low minute volume alarm  0~10 L/min

Continuous pressure alarm   (PEEP+1.5 kPa) over 16s

Suffocation warning   5s~60s no spontaneous ventilation

The maximum limited pressure   <12.5 kPa

Fan error   Show on screen

Oxygen deficit   Show on screen

Working condition

Gas source O2, N2O, AIR

Pressure 280 kPa~600 kPa

Voltage 100~240V

Power frequency 50/60 Hz

GET A FREE QUOTE