YH-1100 ICU Ventilator
这里是标题一h1占位文字
Classification :
Features
1.12.1” TFT screen displays the ventilation parameters, alarm information, and oscillograms, make every operation more easily.
2.Multiple ventilation mode can meet different clinical requirements.( IPPV, A/C, PCV, SIMV,PSV, SPONT/CPAP,SIGH,MANUAL)
3.3 oscillograms for your choice, 2 of them can be displayed on the screen at the same time.
4.Humidifier can heat and wet breathing gas, makes it comfortable for patient to breathe.
5.Rapid oxygen supply,automatically offer high flow rate oxygen within two minutes
6.High temperature resistance breathing circuit is reusable and anti-pollution
Safety
1.14 types of sound and visual alarm information, easier for users to do some error checking and troubleshooting.
2.Built-in oxygen concentration sensor, ensure stable precision of oxygen concentration.
3.Easy to move with four casters, easy to stop with two brakes.
4.Separate design of electronic circuit and gas flow rate keep safe running of ventilator.
5.Compact long life internal battery can provide emergency power,avoid risk of patient.
6.Self-check before operation,eliminate system mistake.
Ventilation Mode
IPPV, A/C, PCV, SIMV, SPONT/CPAP, PSV, SIGH, MANU
Ventilator Parameter
Tidal volume(Vt) 0mL~2000 mL
Frequency (Freq) 1 min~100 min
Oxygen concentration 21 %~100 %
I:E 4:1~1:8
PEEP 0 cmH2O~40 cmH2O
Pressure limit 20 cmH2O~100 cmH2O
Pressure triggering sensitivity (Ptr) -20 cmH2O~0 cmH2O (Based on PEEP)
Flow triggering sensitivity(Ftr) 0.5 cmH2O~30 cmH2O
SIGH 0 (off) 1/100~5/100
Apnea ventilation OFF,5s~60s
Monitoring Parameter
Frequency(Freq) 0 /min~100 /min
Tidal volume(Vt) 0 mL~2500 mL
MV 0 L/min~99 L/min
Airway pressure 0 cmH2O ~ 100 cmH2O
Dynamic lung compliance monitoring 1 mL/cmH2O ~ 1000 mL/cmH2O
Oxygen concentration 15 %~100 %
Alarm and protection
The AC power failure alarm Power failure or no connection
Internal backup battery low voltage alarm ≤ 11.3 ± 0.3 V
No tidal volume No tidal volume within 6 s
High minute volume alarm 5 L/min ~ 99 L/min
Low minute volume alarm 1 L/min ~ 30 L/min
High airway pressure alarm 20 cmH2O~100 cmH2O
Low airway pressure alarm 0 cmH2O~20 cmH2O
High oxygen concentration alarm 19%~100%
Low oxygen concentration alarm 18%~99%
Continuous pressure alarm (PEEP+1.5 cmH2O) over 16s
Suffocation warning 5s~60s
Fan error Show on screen
Oxygen deficit Show on screen
The maximum limited pressure <12.5 kPa
Working condition
Gas source O2, AIR
Pressure 280 kPa~600 kPa
Voltage 220 V ± 22 V
Power frequency 50 Hz ± 1 Hz
Input power
900 VA(With air compressor)
250 VA(Without air compressor)
GET A FREE QUOTE