YH-200A Plus Neonatal CPAP
这里是标题一h1占位文字
Classification :
Neo-CPAP System
Application
1.Continuous Positive Airway Pressure (CPAP) nurtures and supports infant breathing by providing respiratory support throughout the respiratory cycle.
2.Neo-CPAP System maintains the infant's functional residual capacity by helping to prevent airway closure.
3.Neo-CPAP System features an ergonomic and uncluttered touch-screen interface, enabling the clinician to quickly and easy access the current patient status.
4.With the advantages of compact device, small footprint, saving energy and integrated high-performance in operation. This makes it an ideal companion for your smallest patients in the Neonatal Intensive Care Units (NICU), emergency ward, delivery room and other departments.
Trust Point
1.10.4 Inch TFT touch LCD screen displays the ventilation parameters, alarm information and oscillogram.
2.Better work of breathing: Integration breathing circuit design, maximize lung protection and breathing support.
3.Increase clinician's confidence in Neo-CPAP: Multiple parameters monitoring interface, makes every parameter clear, let users know the patient conditions in all aspects.
4.Real time pressure-time oscillogram and high precision O2concentration detection function included, reduces length of stay in hospital.
5.Built-in air and oxygen blender, ensures stable oxygen concentration.
6.Stable and low noise air compressor creates quiet work environment for doctors, can be used for 2-3 Neo-CPAP Systems together.
Ventilation mode
CPAP, NIPPV, HFNC, MANUAL, NCPAP
Parameter
Oxygen concentration
Adjusting range: 21% ~ 100%
Monitoring range: 15% ~ 100%
CPAP
Adjusting range: 0 cmH2O ~ 15 cmH2O
Monitoring range: 0 cmH2O ~ 60 cmH2O
Continuous flow
Adjusting range: 2 L/min ~ 20 L/min
Monitoring range: 0 L/min ~ 20 L/min
Pressure monitoring 0 cmH2O ~ 60 cmH2O
Inhalation pressure 3 cmH2O ~ 15 cmH2O
Exhalation pressure 1 cmH2O ~ 13 cmH2O
Breath Frequency
Adjusting range: 1 /min ~ 100 /min
Monitoring range: 0 ~ 100 /min
Inhalation time
Adjusting range: 0.1 s ~ 15 s
Monitoring range: 0 s - 15 s
Manual ventilation Output flow: 20 L/min ± 5 L/min
Air compressor
Continuous output flow: ≥ 25 L/min
Peak output flow: ≥ 120 L/min
Oscillogram
P-T (pressure – time)
Flow of oxygen, air and mixed gas
Airway pressure oscillogram
Alarm and protection
The AC power failure alarm Power failure or no connection
Internal battery backup low voltage alarm < 11.3 ± 0.3 V
High oxygen concentration alarm 19% ~ 100%
Low oxygen concentration alarm 15% ~ 99%
High Airway pressure alarm 1.5 kPa ~ 6k Pa
Low air input pressure alarm < 200 kPa
Low oxygen input pressure alarm < 200 kPa
Air and oxygen blender alarm
When any gas source lack
When differential pressure > 0.1m Pa
Compression pump alarm Excess temperature
Working condition
Gas source O2, Air
Pressure 280 kPa ~ 600 kPa
Voltage 100 ~ 240 V
Power frequency 50 Hz ± 1 Hz