DM6A Veterinary Anesthesia Machine
这里是标题一h1占位文字
Classification :
Gas supply
Gas supply O2(For medical use)
N2O(For medical use)
Input pressure range 280 kPa~600 kPa
Output pressure of regulator
(gas cylinder supply) less than 440 kPa
Safety valve 750kPa
Negative pressure air source ≥75 L/min
Flow rate
Adjust range of flow meter
O2 0 L/min~10 L/min
N2O 0 L/min~10 L/min
precision(20℃ 101.3kPa, supply pressure 440 kPa)
The precision from 10 % of full scale to full scale is ±10 % of indicated value
The precision lower than 10 % of full scale is class 4.
Linkage control of O2 and N2O flow rate
Adjust O2 concentration to no less than 25 % according to proportion.
Flow rate of oxygen flush valve 25 L/min~ 75 L/min
Power supply
Voltage 100-240 V~
Frequency 50/60HZ
Input power 80VA
Fuse
Main power T2AH 250V
Internal power T4AH 250V
Ventilation mode
IPPV
A/C
SIMV
PCV
Ventilation parameters
Specification
Frequency
Adjusting range 1 bpm ~ 100 bpm
(Under SIMV:1 bpm~ 40 bpm
All mode except SIMV:
VTH 4 bpm~40 bpm
VTL 20bpm~100 bpm)
Allowable error ± 15 %
Inhalation time(I:E)(Tinsp)
Adjusting range VTH 4:1(1:0.25) ~ 1:8
VTL 2:1(1:0.5) ~ 1:8
Tinsp 0.2s~12.0s
Allowable error ± 15 % or ± 0.1 s,
whichever is the greater
Tidal volume(VT)
Adjusting range 0,20 mL ~ 1500 mL
And:
VTH 250 mL ~ 1500 mL
VTL 0,20 mL ~ 300 mL
Minute ventilation(MV)
Max MV
VTH ≥18 L/min
VTL ≥10 L/min
PEEP
Adjusting range 0 kPa ~ 3kPa
Allowable error ±15 % or ±(0.3 kPa+10 % setting valve)whichever is the greater
Pressure trigger(PTr)
Adjusting range -2.0 kPa ~ 0kPa(Under PEEP)
Allowable error ±(0.1 kPa+10 %setting valve)
Flow trigger(FTr)
Adjusting range OFF, 0.5 L/min ~ 30 L/min
Allowable error >3L/min: ±20% ≤3L/min: ±0.6 L/min
Pressure control(Pc)
Adjusting range 0.5kPa ~ 6.0 kPa
Allowable error ± 15 % or ±(0.3 kPa+10 % setting valve)whichever is the greater
Holding time(Inhalation platform)(TIP)
Adjusting range 0 ~ 6s
(0% ~ 50% of inhalation time)
Allowable error ± 15 % or± 0.1 s,whichever is the greater
Sigh
Adjusting range 0, 1/100 ~ 5/100
(Sigh tidal volume is 1.5 to 2 times higher than setting valve of tidal volume)
Apnea ventilation
Adjusting range OFF, 5 s~ 60 s
Max set working pressure(pressure limit range)
Adjusting range 3kPa ~ 10kPa
Allowable error ± 15 % or ±(0.2 kPa+10 % setting valve)whichever is the greater
Max pressure ≤12.5 kPa
explanation:
1 set pressure limit value to produce maximum working pressure( the upper limit of high airway pressure alarm).
2 do not use negative pressure at exhalation.
3 maximum limit pressure is pressure of the safety valve.
Driven gas
Gas supply O2(For medical use)
Input pressure range 280 kPa ~ 600 kPa
Fresh gas compensation
Compensation range 0 ~ 15 L/min
gas composition O2, N2O and anaesthetic
Ventilation parameter Monitoring
The following are monitoring parameters under the environment of body temperature and pressure-saturated.
Minimum exhalation volume maybe monitored is 50 mL/min.
Fig. display:
|
Parameter |
Range |
Resolving power |
Accuracy |
|
Frequency(Freq) |
1 bpm ~ 100 bpm |
1 bpm |
±15 % |
|
Tidal volume(VT) |
0~ 2000 mL |
10 mL |
>100 mL:±20% ≤100 mL:±0.02 L |
|
Minute ventilation(MV) |
0~100L/min |
0.1 L/min |
>1L/min:±20% ≤1L/min:±0.5 L/min |
|
(Ppeak) |
0~100 cmH2O |
1 cmH2O |
±(4%full scale+4%actual reading) |
|
O2 concentration |
15 % ~ 100 % |
0.1 % |
±[ 2.5 %(v/v)+2.5 %(concentration level)] |
|
Compliance |
1~1000mL/ cmH2O |
1mL/ cmH2O |
|
Waveform display:
——Time—airway pressure(under all modes)
——Time—flow(under all modes)
——pressure volume loop(under all modes)
——Time—volume(under all modes)
——Time—CO2(under all modes)
Remarks:The machine cannot record all the adjusting and monitoring values.