Defibrillator
这里是标题一h1占位文字
Classification :
Specification
Display
Display Type: 8 inch high resolution LCD displaySweep Speed:25mm/secInformation: HR, Lead/Pads, On/Off Alarm, SpO2, AED Functions and Prompts, Alarm Se-lection and Limits, Delivered Energy.
Defibrillator
Waveform: BiphasicCharge Time: Less than 7 seconds with a new fully charged battery.Energy Display: Monitor display indicates both selected and delivered energy.Charge Prompt Type: Voice and visual prompts.Electrode lmpedance Measurement Range: 0-250 ohms.
ECG Monitoring
Patient Connection: 5-lead ECG cable, or 3-lead ECG cable, paddles.Lead Selection: Displayed on monitor, paddles, l, l, ll, AVR, AVL, AVF, V.ECG Size: 0.25, 0.5,1,1.5,2, 4 cm/mV diplay on monitor.Heart Rate:20-300BPM.
Heart Rate Alarm: On/Off displayed on monitor, user-selectableSmart Alarms: Beeper/voice prompts indicate shockable rhythm.
Recorder
Paper: 50mm thermal.
Speed:12.5mm/sec,25mm/sec,50mm/sec.User-selectable 6-second delay.
Printing Method: High-resolution, thermal print head.
Print-out Modes: Manual or automatic, user-configurable.
On/Off Control: Front panel and paddle.
Automatic Function: 9-seconds recording initiated by alamm activation or defibrillator chargeor defibrillator discharge.
Battary
Type: Rechargeable, Ni-MH battery, 12V.
Operating Time: For a new, fully charged battery: 60 defibrillator discharges at maximumenergy, or 3 hours minimum of continuous ECG monitoring.Additional parameters will effect operating time with different functions.
AED Mode
AED Function: Auto analyze and charge X3 with programmable auto energy level selection,screen prompts, and voice prompts.
Shockable Rhythms: Ventricular fibrillation with amplitude >=200UV, ventricular tachycardiawith rates >=140bpm,
and QRS complex wave duration >=140ms.
Charge Control: Control on device front panel, press key on paddlePrompts: Voice and visual prompts.
Manual Mode
Energy Selection: Selectable at 2,5,7,10,20,30,50,70,100,150,200,300, 360 joules.Synchronized Mode: Synchronizes defibrillator pulse to patient's R-wave. "SYNC" messagedisplayed on monito
Specification
| Specification | MB_pacer (Optional) |
|---|---|
| Type | VVI |
| Pulse Type | rectangular,constant current |
| Pulse Amplitude | 0 to 180 mA ± 10% or 5 mA (whichever is greater). |
| Pacing Rate | Variable from 30 ppm to 180 ppm ±1.5% (increments or decrements by a value of 2 ppm) |
| Multi-Function Electrode (MFE)Pads | multipurpose defibrillation/pacing electrodes |
| Pause | Pacing pulse frequency reduced by a factor of 4 when activated |
| Refractory Period | NC(VVI demand do not have this specification) |
| Pulse Width | (20+ 1.5 ms) |
| Output Protection | Fully defibrillator protected and isolated |
| Specification | MB_EtCO2 (Optional) |
|---|---|
| Type of sensor | Sidestream |
| Technical principle | Non-dispersive infrared gas analysis NDIR |
| Storage condition | -40 °C to 70°C, <90% RH, non-condensing |
| Operating conditions | 5 °C to 50 °C, 10 to 90% RH, non-condensing |
| Ambient pressure | 55-115kPa |
| Power supply | 5 V±5% (max ripple 200 mVp-p) |
| TDP | Typical value120mA Excursion calibration typical value280mA |
| Range | 0-19.7% (0 –150mmHg,or 0-20kPa) |
| Resolution | 0.1mmHg |
| Accuracy | 0 - 40 mmHg ± 2 mmHg41 - 70 mmHg ± 5% of reading71 - 100 mmHg ± 8% of reading101 - 150 mmHg ± 10% of reading |
| Respiratory rate | 2-150 BPM |
| Respiratory rate measurement accuracy | 1% ±1BPM |
| Specification | MB_NIBP (Optional) |
|---|---|
| Measurement Technology | Oscilloscope OPE measure |
| Parameters Display | Systolic, diastolic and mean pressure |
| Measurement Mode | Manual, Auto and Sequence(Adult Mode) |
| Resolution | 1 mmHg |
| NIBP Range, Accuracy: | Systolic Pressure: 0kPa ~ 34kPa(0mmHg ~ 255mmHg); Accuracy: ±0.5kPa (±4mmHg)Diastolic Pressure: 2kPa ~ 26 kPa(15mmHg ~ 195mmHg); Accuracy: ±0.5kPa (±4mmHg)Mean blood pressure: 2.67kPa~31.3kPa(20mmHg~235mmHg); Accuracy: ±0.5kPa (±4mmHg) |
Specification
| Pressure accuracy | ±0.5kP(±4mmHg) |
|---|---|
| Pressure range | 0kPa(0mmHg)~40kPa(300mmHg) |
| Alarm range can be set | Systolic Pressure: 0kPa~40kPa(0mmHg~300mmHg)Diastolic Pressure: 0kPa~26.7kPa(0mmHg~200mmHg)Mean blood pressure: 0kPa~40.0kPa(0mmHg~300mmHg) |
| Alarm accuracy | Not exceeding the set value ±10% |
| Alarm Time | No more than 12s |
| Specification | SPO2 Moudle (Optional) |
|---|---|
| Measurement Range: | 30 ~100%,±2% between 80%~90%,Others ±5% |
| Alarm Range | User set high limit and low limit |
| Alarm Accuracy | ±10 within setting values. |
| Alarm Time Accuracy | Less than 12 sec. |
GET A FREE QUOTE
RELATED PRODUCTS