HF-210A Warming Blanket
这里是标题一h1占位文字
Classification :
Description:
Warming Blanket utilizes the convection of air to transmit the hot air generated by the main unit to the disposable blanket through the pipeline, so that the heat energy is evenly distributed and transmitted to the patient's body surface covered by the blanket to achieve the eff ect of heat preservation or heating for prevention and treatment of hypothermia in patients. The blanket is made of non-woven fabric and melt-blown cloth delicate material. It is with scientific air outlet arrangement, good temperature transfer eff ect. It is widely used in Operation Room, ICU and other wards. Color LCD screen and quick setting button makes it user friendly.
Features:
8 adjustable wind speed can apply to different patient in different environment
Dual hardware and software temperature protection
More than 10,000 logs for necessary search
With G4 (EN779) grade air filter and plasma air purification function.
16 kinds disposable blankets optional
Detail Spec:
Temperature Setting 32.0 ~41.0°C、95.0 ~105.8°F
Temperature Increment 0.1°C, 0.1°F
Temperature Deviation <±1°C, <±1.8°F
Optional Unit °C & °F
Warming Up Time <4 minutes
Working Mode Continuous working
Operation Time Real time display
Alarm Overheat, low temperature, no operation, fan error, temperature sensor error
Display 3.5-inch color LCD display
Air Volume 8 Level
Natural Wind Function Available
Quick Temperature Setting 4 Temp.
Alarm Volume 8 Level
Mute Time 2 minutes
History Record 10000pcs
Ambient Temperature Function Available
Language Chinese and English
Air Purification Plasma air purification
Blanket 16 kinds disposable blankets optional
Filter G4 (EN779) grade
Power Consumption 850VA
Power Supply AC220V, 50Hz/60Hz
Type of Protection against Electric Shock Class I
Degree of Protection against Electric Shock Applied part Type BF protection against defibrillation discharge eff ects
Operation Environment Temp: +10°C~+40°C; Humidity: 30%~ 75%; Atmospheric pressure: non-condensation, 70KPa~106Kpa
Storage Environment Temperature: -20°C~+55°C; Humidity: 10~ 93%; Atmospheric pressure: non-condensation, 50KPa~106Kpa
Degree of Protection against Ingress of Liquid IPX2
Dimension 260mm X 325mm X 325mm
Weight <6Kg
Overheat Protection Function
Dual hardware and software protection,
real-time temperature monitoring and over 50±5°C hardware forced power off protection