DM6B Veterinary Anesthesia Machine
这里是标题一h1占位文字
Classification :
Gas supply
Gas source: O2(For medical use)
Input pressure range: 280 kPa~600 kPa
Output pressure of regulator
(gas cylinder supply) less than 440 kPa
Safety valve: 780kPa
Negative pressure air source: ≥75 L/min
Flow rate
Adjust range of flow meter
O2 0 L/min~5 L/min
precision (20℃ 101.3kPa,supply pressure 440 kPa)
The precision from 10 % of full scale to full scale is ±10 % of indicated value
The precision lower than 10 % of full scale is class 4.
Flow rate of oxygen flush valve: 25 L/min~ 75 L/min
Output pressure for fresh valve: 30 kPa ~ 45 kPa
Power supply
Voltage: 100-240 V ~
Frequency: 50/60 HZ
Input power: 40VA
Fuse
Main power: T2AH 250V
Internal power: T4AH 250V
Ventilation mode
IPPV
A/C
SIMV
Ventilation parameters
Specification
Frequency
Adjusting range: 1 bpm ~ 100 bpm
(Under SIMV:1 bpm~ 20 bpm All mode except SIMV: 4 bpm~100 bpm)
Allowable error: ± 15 %
Inhalation time (I:E) (Tinsp)
Adjusting range 2:1(1:0.5) ~ 1:6 (0.2s ~ 6.0s)
Allowable error ± 15 % or ± 0.1s, whichever is the greater
Tidal volume (VT)
Adjusting range 0,20 mL ~ 1500 mL
Minute ventilation (MV)
Max MV ≥18 L/min
PEEP(optional)
Adjusting range 0 kPa ~ 3kPa
Allowable error ± 15 % or ±(0.3 kPa+10%setting value), whichever is the greater
Pressure trigger(PTr)
Adjusting range - 2.0 kPa ~ 0kPa(Under PEEP)
Allowable error ±(0.1 kPa+10 %setting value)
Flow trigger(FTr)
Adjusting range OFF, 0.5 L/min ~ 30 L/min
Allowable error >3L/min:±20% ≤3L/min:±0.6 L/min
Holding time(Inhalation platform)(TIP)
Adjusting range 0 ~ 3s(Less than 50% of inhalation time)
Allowable error ± 15 % or± 0.1 s,whichever is the greater
Sigh
Adjusting range 0.1/100 ~ 5/100
Apnea ventilation
Adjusting range OFF, 5 s ~ 60 s
Max set working pressure
pressure limit range 0.3kPa ~ 10kPa
Allowable error ± 15 % or ±(0.2kPa+10%setting value), whichever is the greater
Max pressure ≤12.5 kPa
explanation:
1 set pressure limit value to produce maximum working pressure( the upper limit of high airway pressure alarm).
2 do not use negative pressure at exhalation.
3 maximum limit pressure is pressure of the safety valve.
Driven gas
Gas source O2(For medical use)
Input pressure range 280 kPa~600 kPa
Fresh gas compensation
-
Compensation range 0~15 L/min
-
gas composition O2 and anaesthetic
Ventilation parameter Monitoring
The fllowing are monitoring parameters under the environment of body temperature and pressure-saturated.
Minimum exhalation volume maybe monitored is 50 mL/min.
display:
|
Parameter |
Range |
Resolving power |
Accuracy |
|
Frequency(Freq) |
0 bpm ~ 100 bpm |
1 bpm |
±15 % |
|
Tidal volume(VT) |
0~ 2000 mL |
10 mL |
>100 mL , ±20% ≤100 mL, ±0.02 L |
|
Minute ventilation(MV) |
0~100L/min |
0.1 L/min |
>1L/min, ±20% ≤1L/min, ±0.5 L/min |
|
(Ppeak) |
0~100 cmH2O |
1 cmH2O |
±(4%full scale+4%actual reading) |
|
O2 concentration(optional) |
15 % ~ 100 % |
0.1 % |
±[ 2.5 %(v/v)+2.5 %(concentration level)] |
|
Compliance |
1~1000mL/ cmH2O |
1mL/ cmH2O |
|
Waveform display:
Time- airway pressure(under all modes)
Time—flow(under all modes)
pressure volume loop(all modes)
Remarks:The machine cannot record all the adjusting and monitoring values.