• YH-B3000 Infant Incubator
YH-B3000 Infant Incubator

YH-B3000 Infant Incubator

这里是标题一h1占位文字

Servo Control Mode: air temperature, skin temperature, display humidity; Parameter Displaying: set temperature,air temperature, skin temperature, internal humidity, heat power rate; Audible & Visual Alarms: power failure, sensor failure,

Product feature:

Servo Control Mode: air temperature, skin temperature, display humidity;

Parameter Displaying: set temperature,air temperature, skin

temperature, internal humidity, heat power rate;

Audible & Visual Alarms: power failure, sensor failure,

higher & lower temp, air & skin temperature override,

fan failure, system failure;

>37℃ Preset Mode: power failure reset memory, alarm silence,

keypad lock, self-test function;

Hood with dual slope; 6 hand-ports;

RS-232 Output Connector (use for network system);

IV poles;

Protective guard rail;

Storage cabinets;

Can option humidity control.

Specifications of hood

Mattress to Hood: 450 mm

Observation side of hood: Two incline

Operate port: 6 (2 Iris ports)

Tubing grommets: 4

Front access panel: 826×310 mm

Mattress Tray Size: 630×360 mm

Infant bed tilt adjusting ±10° step-less

Technical Specifications 

Power requirement ≤400VA

Temperature control mode:

Air and skin servo control

Air Temp Control Range: 20.0℃-37.0℃

Temperature override mode to: 37.1℃-39.0℃

Skin Temp Control Range: 20.0℃-37.0℃

Temperature override mode to: 37.1℃-38.0℃

Skin temp sensor accuracy: ≤0.3℃

Humidity display range: 5%-99%

Temperature fluctuation: ≤0.5℃

Mattress temperature uniformity: ≤0.8℃

Temperature rise time: ≤45 min

Noise level within hood environment: ≤55dB

Alarms

Power failure alarm

Fan failure alarm

Sensor failure alarm

High air temperature alarm: +3.0℃

Low air temperature alarm: -3.0℃

High skin temperature alarm: +1.0℃

Low skin temperature alarm: -1.0℃

Over air temperature alarm: ≤38.0℃; ≤40.0℃

Over skin temperature alarm: ≤38.0℃; ≤39.0℃

System failure alarm

Alarm silence function

Operational condition

Environment temperature: +20℃-+30℃

Environment relative humidity: 30%-75%

Environment air velocity of flow: <0.3m/s

Others

Mattress tray Max. load 5kg

Transfusion pole Max. load 2kg

Cabinet Max. load 6kg