• YH-B1000 Infant Incubator
YH-B1000 Infant Incubator

YH-B1000 Infant Incubator

这里是标题一h1占位文字

1.Servo Control Mode: air temperature; 2.Parameter Displaying: set temperature,air temperature, heat power rate; 3.Audible & Visual Alarms: power failure, sensor failure, higher & lower temp, air temperature override,fan failure, system failure; 4.>37℃ Preset Mode: power failure reset memory, alarm silence, keypad lock, self-test function;

Product feature:

1.Servo Control Mode: air temperature;

2.Parameter Displaying: set temperature,air temperature, heat power rate;

3.Audible & Visual Alarms: power failure, sensor failure, higher & lower temp, air temperature override,fan failure, system failure;

4.>37℃ Preset Mode: power failure reset memory, alarm silence, keypad lock, self-test function;

5.Hood with dual slope; 6 hand-ports;

6.RS-232 Output Connector (use for network system);

7.IV poles;

8.Protective guard rail;

9.Storage cabinets.

Specifications of hood

Mattress to Hood: 450 mm

Observation side of hood: Two incline

Operate port: 6 (2 Iris ports)

Tubing grommets: 4

Front access panel: 826×310 mm

Mattress Tray Size: 630×360 mm

Infant bed tilt adjustable ±10° step-less

Technical Specifications 

Power requirement ≤400VA

Temperature control mode: Air servo control

Air Temp Control Range: 20.0℃-37.0℃

Temperature override mode to: 37.1℃-39.0℃

Temperature fluctuation: ≤0.5℃

Mattress temperature uniformity: ≤0.8℃

Temperature rise time: ≤45 min

Noise level within hood environment: ≤55dB

Alarms 

Power failure alarm

Fan failure alarm

Sensor failure alarm

High air temperature alarm: +3.0℃

Low air temperature alarm: -3.0℃

Over air temperature alarm: ≤38.0℃; ≤40.0℃

System failure alarm

Alarm silence function

Operational condition 

Environment temperature: +20℃-+30℃

Environment relative humidity: 30%℃-75%

Environment air velocity of flow: <0.3m/s

Others 

Mattress tray Max. load 5kg

Transfusion pole Max. load 2kg

Cabinet Max. load 6kg